Abstract
Animal magnetic resonance imaging (MRI) systems typically deliver superior imaging performance over conventional human MRI systems, making them a prevailing instrument in preclinical research. It is challenging to achieve the high performance of these animal MRI systems, due to the multifaceted nature of the various system components and the complexity of integration debugging. This work described the design, fabrication, measurement and integration of a 7 T animal MRI system, which exhibits several performance highlights. Both the magnet and gradient assembly adopted an ultra-shielding strategy, facilitating ease of system installation, maintenance and debugging. The main magnetic field exhibits acceptable homogeneity and stability, and the gradient coil is mechanically reliable thanks to zero-force control. The animal MRI system underwent debugging using proprietary imaging software to generate images of phantoms, fruits and organisms. Further research investigation will be performed to promote more scientific outputs with enhanced functional capabilities.
Animal magnetic resonance imaging (MRI) systems typically incorporate ultra-high-field magnets and powerful gradient assemblies, which generate very high signal-to-noise (SNR) ratios and image resolution to support fundamental research in medicine and life sciences1,2,3. To develop these ultrahigh-field animal MRI systems, the challenging aspects include but are not limited to, generating an ultrahigh magnetic field strength at the central area with acceptable field homogeneity and simultaneously shielding the strong stray field; and engineering a strong gradient magnetic field with minimal magnetic and mechanical coupling on the magnet. In contrast, generating the radio-frequency (RF) magnetic field is relatively straightforward compared to the whole-body human counterpart, due to the smaller imaging size and the lack of need for specific RF wavelength considerations4,5,6. The electrical and electronic modules include a spectrometer, gradient power supply and amplifier, and RF power supply and amplifier. The system integration and debugging depend on robust software and pulse sequences.
The research of animal MRI system has been an important topic. In superconducting magnets, apart from the inherent reliability parameters and assembly units, such as safe current margin, controllable stress level, reliable quench protection circuit7,8,9,10,11, superconducting joints, switches and current leads12,13, the most critical external parameters include a homogeneous distribution over the imaging area, 5 Gauss limit, bore size and magnet dimension. Different from permanent magnet and resistive magnet, cryogenic heat transfer is a crucial aspect of the superconducting magnet technology14, the stable operation of the superconducting magnet needs a stable ultralow-temperature environment. After current excitation, shimming is essential to compensate for the magnetic field inhomogeneity caused by the engineering development process15,16.
Ultrahigh gradient magnetic fields require large coil currents, which induce large Lorentz forces under an ultrahigh static magnetic field. Unbalanced Lorentz force of the gradient coils can lead to vibrations in the MRI system, posing a risk to the magnet’s safety and potentially inducing magnetic field fluctuations17,18. In addition, shielding ultrahigh gradient magnetic fields is critical to reduce secondary magnetic field interference on the gradient magnetic field linearity19. Improper design of the shielded gradient coil can result in large eddy current artifacts20. In addition to the three-axis gradient coils, an animal gradient assembly usually includes room-temperature shim coils and passive shimming tunnels to offset magnetic field non-homogeneous harmonics in the magnetic field, as well as an efficient cooling path to dissipate the heat generated by the coils.
Animal MRI systems have been promoted by several companies for many years. In the market, Bruker animal MRI systems are currently in a dominant position21. Some of them have been developed into hybrid instrument with other imaging modalities22 or therapeutic devices23. There is also successful integration of animal MRI system using a dedicated animal superconducting magnet and human imaging platform24. With small gradient assembly insert, a background whole-body MRI system can be transformed into a provisional high resolution animal imaging system25. While these systems have significantly contributed to preclinical research and scientific studies, most works concentrated on RF coil or imaging methods improvement and the literature lacks detailed reports on the latest advanced update of the hardware performance such as superconducting magnet and gradient assembly.
This work highlights the technological advancements of a proprietary ultrahigh-field animal MRI system, mainly focusing on the superconducting magnet and gradient assembly, and the imaging evaluations were also conducted. The super-shielding and mechanical decoupling designs of the superconducting magnet and gradient assembly contribute the critical technical features of animal MRI system.
Results
Superconducting magnet
The actively-shielded 7 T animal MRI superconducting magnet coil structure is shown in Fig. 1a, which was wound by NbTi superconducting wire. The adjusting coils consist of four paratactic solenoid coils. These coils contribute magnetic field strength to the primary coils and significantly reduce the inhomogeneous harmonic components of the central magnetic field distribution, thus forming a magnetic field homogeneous region with 10 ppm homogeneity as shown in Fig. 1b. Unlike the conventional double-coil pattern, the shielding coils include an additional coil in the center. The three-coil shielding pattern creates a compact 5 Gauss line range within ±2.95 m at the axial direction and ±1.85 m at the radial direction as shown in Fig. 1c. The commercial counterpart system has a 5 Gauss range of ±3 m at the axial direction and ±2 m at the radial direction. Especially in the central section of the 5 Gauss range, the minimum radius is less than 1.4 m, which supports several safe operations such as plugging and unplugging the current lead, liquid helium infusion, cold head maintenance, etc. The coil patterns and magnetic field contours are exactly output based on the magnet design parameters displayed in Table 1, including the individual coil dimension with left axial position z1, right axial position z2, inner radius r1, outer radius r2 and corresponding current density. The critical performance parameters of the superconducting magnet design are listed in Table 2.
Fig. 1: Design of a 7 T animal MRI superconducting magnet.
figure 1
a magnet coils, b central magnetic field homogeneity distribution with ±5 ppm contour and c 5 Gauss line range surrounding the magnet.
Full size image
Table 1 Individual coil dimension and corresponding current density
Full size table
Table 2 Design parameters of the superconducting magnet
Full size table
The superconducting shim coils design is shown in Fig. 2a, which has a layer order from inside to outside with Z2, Z1, Z3 by solenoid winding and then X, Y, ZX, ZY, X2-Y2, XY by saddle winding. We generally implemented the maximum sensitivity strategy to design the shim coils, thus it could cope with a worst inhomogeneous situation. The decoupling optimization design of shim coil Z2 makes it an alternating positive and negative seven sub-coils layout, which is much different from the conventional four-coil pattern and the mutual inductance of the magnet and shim coil is reduced from 3.9673 H to only 0.0041 H. All the coils have an identical magnetic field deviation ±1% over the central spherical area, whose magnetic field patterns are plotted in Fig. 2b. Some basic design parameters of the superconducting shim coils are listed in Table 3.
Fig. 2: Superconducting shim coils design and corresponding magnetic field patterns.
figure 2
a–i shim coil of Z2 (decoupled), Z1, Z3, X, Y, ZX, ZY, X2-Y2, XY, respectively and j–r magnetic field patterns of corresponding shim coils.
Full size image
Table 3 Superconducting shim coil parameters
Full size table
An illustration of the cryogenic system is shown in Fig. 3. All the superconducting coils including the magnet coils and shim coils are encapsulated in the cryogenic system in order to working at ultralow-temperature environment. The cryogenic system is consisted of a liquid helium vessel, cold shield, vacuum container and cold head. The two-stage cold head connects to the cold shield at the first stage to keep a temperature of around 50 K and suspends at the second stage into the liquid helium vessel to condense the vapored helium gas back into the liquid state. The inner wall of the vacuum container formulates the warm bore of the superconducting magnet with a diameter of 300 mm.
Fig. 3: Cryogenic system of the superconducting magnet.
figure 3
The cryogenic system is consited of a liquid helium vessel, cold shield, vacuum container and cold head. The superconducting magnet coil and shim coil are encapsulated in the liquid helium vessel.
Full size image
Gradient assembly
The gradient coils and room-temperature shim coils design is shown in Fig. 4. The gradient magnetic field strength is 200 mT/m with a magnetic field deviation ±2.5% over the homogeneous area. By minimizing the Lorentz force during the gradient coil design, the ultimate force at the designated gradient magnetic field strength within the real main magnetic field environment was controlled to be nearly zero, with a residual value of less than 0.1 N. By implementing the ultra-shielded optimization strategy, namely the stray magnetic field strength was reduced as much as possible through repeated iterations of the gradient coil design and simultaneously considered the engineering fabrication feasibility, the maximum stray field intensity for the gradient coils was minimized to 4 Gauss. This was realized with a minimum wire spacing of 3.9 mm for the X gradient coil, 4.0 mm for the Y gradient coil and 4.5 mm for the Z gradient coil. The performances of the gradient coils are listed in Table 4. For the gradient coil slew rate, it depends on the voltage, inductance and also influenced by the resistance. The equation for slew rate calculation is S = G(U-IR)/IL, where G is the gradient magnetic field strength, U is the voltage, I is the operating current, R is the resistance and L is the inductance. Assuming that the voltage is 500 V, by substituting the parameters in Table 4, the slew rates for the Z, X, and Y gradient coils are 503.9357 T/m/s, 1070.3454 T/m/s and 929.9312 T/m/s, respectively.
Fig. 4: Gradient coils and room-temperature shim coils design.
figure 4
a–c actively-shielded Z, X and Y gradient coils, respectively, d–g zonal shim coils Z0, Z2, Z3 and Z4, respectively and h–k tesseral shim coils ZX, ZY, X2-Y2 and XY, respectively.
Full size image
Table 4 Gradient coil performances
Full size table
By implementing the maximizing sensitivity optimization strategy, we set the minimum wire spacing for zonal shim coils Z0, Z2, Z3, and Z4 as 3.0 mm, 3.0 mm, 3.0 mm and 3.1 mm, respectively, and 4.0 mm, 4.1 mm, 4.2 mm and 4.2 mm for the tesseral shim coils ZX, ZY, X2-Y2, XY. The transverse gradient and tesseral shim coils were manufactured using copper cutting techniques, while the axial gradient and zonal shim coils were fabricated using solenoid winding techniques. The performances of the room-temperature shim coils are detailed in Table 5.
Table 5 Room-temperature shim coil performances
Full size table
The inner diameter of the gradient assembly is 173.4 mm. The gradient coils own high efficiency amounting to near 1 mT/m/A due to a long coil design that has copper layer amoungting to 1 m. The engraving technique on solid copper plate ensures a low resistance26. A dual-path cooling system supports the effective dissipation of Joule heat.
Magnet measurement and shimming
The superconducting magnet was excited to 7.02 T with magnetic field locking by a nuclear magnetic resonance (NMR) probe, as shown in Fig. 5a. After more than 7 hours of continuous sampling at a fixed position within the magnet warm bore, the decay rate of the magnetic field strength was measured to be 0.0494 ppm/h, which is stable enough for MRI application. The linear fitting calculation of the magnetic field decay rate is illustrated in Fig. 5b. Afterwards, the superconducting switches of the shim coils were individually heated to eliminate the induced currents during magnet excitation. This is necessary to evaluate the magnetic field homogeneity by removing the magnetic field interference caused by induced currents. The initial magnetic field homogeneity was then measured over a 130 mm spherical surface with a spatial positioning device, as shown in Fig. 5c, resulting in a peak-peak homogeneity of 96.32 ppm.
Fig. 5: Magnetic field measurement of the fabricated superconducting magnet.
figure 5
a magnetic field locking by an NMR probe, b magnetic field decay rate evaluation and c magnetic field distribution over a 130 mm spherical surface.
Full size image
The magnet was shimmed firstly by superconducting shim coils and then by room-temperature iron pieces. The current input of the superconducting shim coils was calculated by a harmonic magnetic field fitting, and the corresponding currents were decomposed as 0.3978 A for Z1, –0.2359 A for X, –0.3978 A for Y, 6.0777 A for Z2, 1.3554 A for ZX, –0.3609 A for ZY, 0.1955 A for X2-Y2, –1.4538 A for XY and 0.2828 A for Z3, respectively. After the coils were excited and closed, the magnetic field homogeneity over the 130 mm spherical surface was improved to be 24.02 ppm. Subsequently, two-round passive shimming operations were conducted with separate odd and even shim trays. Two types of iron pieces were used, both having identical cross sections 24.95 mm × 24.95 mm. However, they have different thicknesses and saturation magnetization. The odd shimming operation used iron pieces with a thickness of 0.3 mm and saturation magnetization of 2.24 T. In contrast, the even shimming operation employed iron pieces with a thinner thickness of 0.05 mm and a lower saturation magnetization of 0.85 T. After passive shimming, the magnetic field homogeneity was further improved to be 7.66 ppm. The shimming results are illustrated in Fig. 6. The other homogeneity values over different imaging areas can be calculated by a harmonic decomposition method. That is 2.54 ppm over a 100 mm spherical surface, 1.57 ppm for 80 mm and 0.80 ppm for 50 mm, respectively.
Fig. 6: Superconducting and passive shimming results.
figure 6
a current input on the superconducting shim coils, b iron pieces distribution in the shim slots, c magnetic field distribution over the 130 mm spherical surface after superconducting shimming and d magnetic field distribution over the 130 mm spherical surface after passive shimming.
Full size image
MRI system integration
Due to the effective Lorentz force minimization, the force-free gradient assembly can be operated inside the magnet warm bore without any fixation. Comparatively, conventional gradient assembly has to be extended to the magnet flange and dedicatedly installed. Figure 7a shows the gradient assembly placed in the warm bore with inserted passive shim trays. The fabricated gradient assembly is shown in Fig. 7b and an in-house passively-shielded saddle RF coil27 with inner bore diameter 80 mm is depicted in Fig. 7c. The RF transmit coil is resonated at 299 MHz with experimental measurement. The RF frontend box28 connected to the RF coil is displayed in Fig. 7d and the integration of the spectrometer, gradient power supply and amplifier, RF power supply and amplifier is presented in Fig. 7e. The RF coil was tuned and matched when conducting the imaging operation. The spectrometer has eight transceiver channels with frequency range 292 –299 MHz.
Fig. 7: Integration of the MRI system.
figure 7
a Gradient assembly and RF coil in the magnet warm bore, b fabricated gradient assembly, c RF coil for testing, d RF frontend connection and e integration of the console and power system.
Full size image
Debugging and imaging
A self-developed software package was used to debug the MRI system. The software contains routine pulse sequences, and also supports fat inhibition, water inhibition, dynamic enhanced scanning, motion artifact inhibition, diffusion imaging and angiography imaging. The initial trials of the imaging operation were based on the spin echo (SE) sequence, although there are more advanced sequences in the package. Both phantom and orange were sampled to test the instrument performance and the corresponding images are shown in Fig. 8. The diameter of the phantom is around 30 mm and the diameter of the orange is around 70 mm. The number of average is two, field of view (FOV) is around 100 mm× 100 mm, image pixel is 256 × 256, resolution is around 390 μm, and bandwidth is 25 kHz. The phantom fixed in the RF bore is shown in Fig. 8a and its images are displayed in Fig. 8b with a signal-to-noise (SNR) ratio of 167.6. Due to the limitations of RF bore size, the orange sit snugly against the inner bore where the B1 field is highly inhomogeneous and thus its edges have some dark shadows, which are shown in Fig. 8c and Fig. 8d, respectively.
Fig. 8: Experimental test of the instrument performance.
figure 8
a a phantom in the RF coil, b phantom image, c an orange in the RF coil and d orange image.
Full size image
The debugging processes included determining the center frequency, testing the nuclear magnetic resonance spectrum, selecting pulse sequence, setting FOV, slice thickness and sampling matrix, etc. After the debugging, a rat experiment using SE sequence with time of echo (TE) 10 ms and time of repetition (TR) 50 ms was conducted, and the diameter of the rat head is around 40 mm. The FOV is around 100 mm × 100 mm, acquisition matrix size is 256 × 256, and the slice thickness is 1 mm. To ensure safe and reliable experimental operations, the rat was anesthetized. The SE images are shown in Fig. 9. The physiological structure of the rat can be revealed from these images, and minimal artifacts and external noise interference were observed.
Fig. 9: Biological experiment on the MRI system.
figure 9
a an anesthetized rat fixation on a plate, b a rat in the RF coil, c inserting the RF coil into the bore of the gradient assembly and (d, e) rat images from different slices and orientations.
Full size image
Discussion
A 7 T animal MRI system was developed, including all the hardware components in the entire machine. During the development of critical components, some distinctive technologies were employed. The magnet has a superior shielding performance, characterized by a significantly reduced 5 Gauss line range compared to its commercial alternatives. Through the combined application of superconducting shimming and passive shimming, the ultimate magnetic field homogeneity was achieved at 7.66 ppm within a 130 mm spherical volume. The magnetic stability of the closed-loop magnet system reached 0.0494 ppm/h. These homogeneity and stability metrics are sufficient for standard MRI operations. In the design of gradient coils, a strategy to minimize the maximum stray field was implemented to ensure gradient assembly experiences non-observable eddy current interference during experimental measurements. Meanwhile, the minimization of Lorentz force resulted in residual forces less than 0.1 N at full gradient magnetic field strength, thus creating safe and convenient mechanical installation conditions for the gradient assembly.
Using our in-house developed RF coil, software, and console integration platform, we successfully acquired magnetic resonance images of phantoms, fruits and organisms. Although there are many advanced methods to improve the imaging functions, we mostly focused on hardware development so as to providing a platform for animal model scientific research. With the advanced hardware technological update, it is feasible to develop more specialized magnetic resonance function.
Methods
A compensating field method was used to homogenize the central magnetic field distribution at a diameter of spherical volume (DSV) of 130 mm. The magnet was actively shielded, achieving a very compact range of 5 Gauss. The magnet coil structure was configured into four layers: two inner nested primary coils, an outer adjusting coil layer and an outermost shielding coil layer. An inverse electromagnetic optimization technique was applied to simultaneously achieve the magnetic field target in both the central homogeneous region and the surrounding shielding area29.
A total of nine coils were designed, including Z, X, Y, Z2, ZX, ZY, X2-Y2, XY, and Z3. The Z2 coil was specially considered with inductance decoupling from the magnet to mitigate quench damage30. To minimize winding deviation, the tesseral shim coils were designed with a quasi-saddle geometry31. Each superconducting shim coil is connected in series, equipped with independent superconducting switches and current lead, decoupled from the magnet.
The gradient assembly was designed with a force minimization to decouple the mechanical interaction between the gradient assembly and the magnet32. For the mitigation of magnetic coupling between the gradient coils and the magnet, a strong shielding strategy was implemented. The shielding strategy was realized by a series of design trials, each with progressively decreasing constraints on stray field intensity. The criterion used to determine the limiting factor for stray field intensity is the minimum wire spacing. This ensures the feasibility of fabricating the coil without local filamentary wires that could potentially result in a burning point. In this work, we considered the minimum wire spacing for the gradient coils to be approximately 4 mm. With this criterion, we finally found the minimum stray field strength constraint for the gradient coil design. To determine the sensitivities of room-temperature shim coils, we tried to achieve the highest possible number of turns until reaching the minimum wire spacing of around 3 mm for zonal coils and 4 mm for tesseral coils.
The RF coil is a shielded saddle coil that resonates at around 7 T, and the software is a self-developed one with routine pulse sequences. The console, power supply and amplifiers were commercially acquired and self-assembled. All these components were integrated into a complete animal MRI system.
Data availability
No datasets were generated or analysed during the current study.
Code availability
There is no original code in this work.
References
Benveniste, H. & Blackband, S. MR microscopy and high resolution small animal MRI: applications in neuroscience research. Prog. Neurobiol. 67, 393–420 (2002).
ArticlePubMedGoogle Scholar
Hoyer, C., Gass, N., Weber-Fahr, W. & Sartorius, A. Advantages and challenges of small animal magnetic resonance imaging as a Translational Tool,. Neuropsychobiology 69, 187–201 (2014).
ArticleCASPubMedGoogle Scholar
Kim, S.-G. & Ugurbil, K. High-resolution functional magnetic resonance imaging of the animal brain. Methods 30, 28–41 (2003).
ArticleCASPubMedGoogle Scholar
Regatte, R. R. & Schweitzer, M. E. Ultra-high-field MRI of the musculoskeletal system at 7.0T. J. Magn. Reson. Imaging 25, 262–269 (2007).
ArticlePubMedGoogle Scholar
Ladd, M. E. et al. Pros and cons of ultra-high-field MRI/MRS for human application. Prog. Nucl. Magn. Reson. Spectrosc. 109, 1–50 (2018).
ArticleCASPubMedGoogle Scholar
Williams, S. N., McElhinney, P. & Gunamony, S. Ultra-high field MRI: parallel-transmit arrays and RF pulse design. Phys. Med. Biol. 68, 02TR02 (2023).
ArticleGoogle Scholar
Wang, Y. et al. Actively-shielded superconducting magnet design of a large-Bore 7 T animal MRI scanner. IEEE Trans. Appl. Superconductivity 30, 1–4 (2020).
Google Scholar
Chen, J. & Jiang, X. Stress analysis of a 7 T actively shielded superconducting magnet for animal MRI. IEEE Trans. Appl. Superconductivity 22, 4903104 (2012).
ArticleGoogle Scholar
Wang, Y. et al. Design, fabrication and measurement of a 3 T cryogen-free animal MRI superconducting magnet. IEEE Trans. Appl. Superconductivity 34, 1–4 (2024).
Google Scholar
Chen, J. & Jiang, X. Protection circuit and quench simulation of a 7 T superconducting magnet for animal MRI. IEEE Trans. Appl. Superconductivity 23, 4701104 (2013).
ArticleGoogle Scholar
Li, Y. et al. A passive quench protection design for the 9.4 T MRI superconducting magnet. IEEE Trans. Appl. Superconductivity 24, 1–5 (2014).
Google Scholar
Liu, S., Jiang, X., Chai, G. & Chen, J. Superconducting joint and persistent current switch for a 7-T animal MRI magnet. IEEE Trans. Appl. Superconductivity 23, 4400504 (2013).
ArticleGoogle Scholar
Manso Jimeno, M., Vaughan, J. T. & Geethanath, S. Superconducting magnet designs and MRI accessibility: A review. NMR Biomedicine 36, e4921 (2023).
ArticleCASGoogle Scholar
Bredy, P. et al. Cryogenics around the 11.7 T MRI Iseult magnet. IEEE Trans. Appl. Superconductivity 20, 2066–2069 (2010).
ArticleCASGoogle Scholar
Haskell, M. W., Nielsen, J.-F. & Noll, D. C. Off-resonance artifact correction for MRI: A review. NMR Biomedicine 36, e4867 (2023).
ArticleGoogle Scholar
Gach, H. M., Curcuru, A. N., Mutic, S. & Kim, T. B0 field homogeneity recommendations, specifications, and measurement units for MRI in radiation therapy. Med. Phys. 47, 4101–4114 (2020).
ArticlePubMedGoogle Scholar
Sakhr, J. & Chronik, B. A. Vibrational response of a MRI gradient coil cylinder to time-harmonic Lorentz-force excitations: An exact linear elastodynamic model for shielded longitudinal gradient coils. Appl. Math. Model. 74, 350–372 (2019).
ArticleGoogle Scholar
Hiltunen, J. et al. Quantification of mechanical vibration during diffusion tensor imaging at 3 T. NeuroImage 32, 93–103 (2006).
ArticlePubMedGoogle Scholar
Wang, Y. et al. Gradient coil design with enhanced shielding constraint for a cryogen-free superconducting MRI system. Magn. Reson. Lett. 4, 100086 (2024).
ArticleCASGoogle Scholar
Graham, M. S., Drobnjak, I. & Zhang, H. Realistic simulation of artefacts in diffusion MRI for validating post-processing correction techniques. NeuroImage 125, 1079–1094 (2016).
ArticlePubMedGoogle Scholar
Waterton, J. et al. Repeatability and reproducibility of longitudinal relaxation rate in 12 small-animal MRI systems. Magn. Reson. Imaging 59, 121–129 (2019).
ArticlePubMedPubMed CentralGoogle Scholar
Thiessen, J. D. et al. MR-compatibility of a high-resolution small animal PET insert operating inside a 7 T MRI. Phys. Med. Biol. 61, 7934 (2016).
ArticleCASPubMedGoogle Scholar
Zhang, X. et al. Evaluation of a developed MRI-guided focused ultrasound system in 7 T small animal MRI and proof-of-concept in a prostate cancer Xenograft model to improve radiation therapy. Cells 12, 1–21 (2023).
Google Scholar
Felder, J., Celik, A. A., Choi, C.-H., Schwan, S. & Shah, N. J. 9.4 T small animal MRI using clinical components for direct translational studies. J. Transl. Med. 15, 264 (2017).
ArticlePubMedPubMed CentralGoogle Scholar
Herrmann, K. et al. Possibilities and limitations for high resolution small animal MRI on a clinical whole-body 3T scanner. Magn. Reson. Mater. Phys., Biol. Med. 25, 233–244 (2012).
ArticleGoogle Scholar
Handler, W. et al. Design and construction of a gradient coil for high resolution marmoset imaging. Biomed. Phys. Eng. Express 6, 045022 (2020).
ArticlePubMedGoogle Scholar
Parsa, J. & Mohammadzadeh, M. Design and fabrication of a new multi-loop saddle coil for 1.5 T MRI. Rev. Sci. Instrum. 90, 114707 (2019).
ArticleCASPubMedGoogle Scholar
Vaughan, J. T., Haupt, D. N., Noa, P. J., Vaughn, J. M. & Pohost, G. M. RF front end for a 4.1 Tesla clinical NMR spectrometer. IEEE Trans. Nucl. Sci. 42, 1333–1337 (1995).
ArticleGoogle Scholar
Wang, Y. et al. Actively-shielded ultrahigh field MRI/NMR superconducting magnet design. Superconductor Sci. Technol. 35, 014001 (2022).
ArticleCASGoogle Scholar
Zhu, X. et al. The design of decoupled even-order zonal superconducting shim coils for a 9.4 T whole-body MRI. IEEE Trans. Appl. Superconductivity 26, 1–8 (2016).
CASGoogle Scholar
Wang, Y., Wang, Q., Qu, H., Liu, Y. & Liu, F. Tesseral superconducting shim coil design with quasi-saddle geometry for use in high-field magnet system. Rev. Sci. Instrum. 90, 094705 (2019).
ArticlePubMedGoogle Scholar
Wang, Y. et al. A dichotomization winding scheme on a novel asymmetric head gradient coil design with an improved force and torque balance,. IEEE Trans. Biomed. Eng. 66, 3339–3345 (2019).
PubMedGoogle Scholar
Download references
Acknowledgements
This work is supported by the National Natural Science Foundation of China (Grant No. 52293423 and Grant No.52277031) and the International Partnership Program of Chinese Academy of Sciences (Grant No. 182111KYSB20210014).
Author information
Authors and Affiliations
Institute of Electrical Engineering, Chinese Academy of Sciences, Beijing, 100190, China
Yaohui Wang, Pengfei Wu, Wenhui Yang & Qiuliang Wang
School of Electronic, Electrical and Communication Engineering, University of Chinese Academy of Sciences, Beijing, 100049, China
Yaohui Wang, Pengfei Wu, Wenhui Yang & Qiuliang Wang
School of Electrical and Electronic Engineering, Huazhong University of Science and Technology, Wuhan, 430074, China
Guyue Zhou & Haoran Chen
School of Electrical Engineering and Computer Science, The University of Queensland, Brisbane, QLD, 4072, Australia
Feng Liu
Authors
Yaohui Wang
View author publications
You can also search for this author in PubMedGoogle Scholar
2. Guyue Zhou
View author publications
You can also search for this author in PubMedGoogle Scholar
3. Haoran Chen
View author publications
You can also search for this author in PubMedGoogle Scholar
4. Pengfei Wu
View author publications
You can also search for this author in PubMedGoogle Scholar
5. Wenhui Yang
View author publications
You can also search for this author in PubMedGoogle Scholar
6. Feng Liu
View author publications
You can also search for this author in PubMedGoogle Scholar
7. Qiuliang Wang
View author publications
You can also search for this author in PubMedGoogle Scholar
Contributions
Y. Wang: magnet coil and gradient assembly design, measurement, shimming, system integration, manuscript writing. G. Zhou, H. Chen and P. Wu: shimming, debugging. W. Yang: imaging software and debugging. F. Liu: design methodology, manuscript editing. Q. Wang: research grants, technical guidance, magnet design, system integration, manuscript editing.
Corresponding authors
Correspondence to Yaohui Wang or Qiuliang Wang.
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
Reprints and permissions
About this article
Check for updates. Verify currency and authenticity via CrossMark
Cite this article
Wang, Y., Zhou, G., Chen, H. et al. Ultrahigh-field animal MRI system with advanced technological update. npj Imaging 2, 54 (2024). https://doi.org/10.1038/s44303-024-00060-0
Download citation
Received:22 June 2024
Accepted:19 November 2024
Published:11 December 2024
DOI:https://doi.org/10.1038/s44303-024-00060-0
Share this article
Anyone you share the following link with will be able to read this content:
Get shareable link
Sorry, a shareable link is not currently available for this article.
Copy to clipboard
Provided by the Springer Nature SharedIt content-sharing initiative